11.3.2. Table 202

Guidance and manufacturer’s declaration – electromagnetic immunity – for all EQUIPMENT
and SYSTEMS (see 6.8.3.201 a) 6))

Guidance and manufacturer´s declaration – electromagnetic immunity

The PENTA is intended for use in the electromagnetic environment specified below. The customer
or the user of the PENTA should assure that it is used in such an environment.

Immunity test

IEC 60601
test level

Compliance level

Electromagnetic environment -
guidance

Electrostatic
discharge
(ESD)

IEC 61000-4-2

± 6 kV contact

± 8 kV air




± 6 kV contact

± 8 kV air


Floors should be wood, concrete
or
ceramic tile. If floors are covered
with
synthetic material, the relative
humidity
should be at least 30 %.

Electrical fast
transient / burst

IEC 61000-4-4



± 2 kV for power
supply lines

± 1 kV for
input/output
lines




± 2 kV for power
supply lines

± 1 kV for input/output
lines

Mains power quality should be
that of a
typical commercial or hospital
environment.

Surge

IEC 61000-4-5


± 1 kV differential
mode

± 2 kV common
mode





± 1 kV differential
mode

± 2 kV common mode


Mains power quality should be
that of a
typical commercial or hospital
environment.

Voltage dips,
short
interruptions
and
voltage
variations
on power
supply
input lines

IEC 61000-4-11

< 5 % U

T

(>95 % dip in U

T

)

for 0,5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

< 5 % U

T

(>95 % dip in U

T

)

for 5 sec

< 5 % U

T

(>95 % dip in U

T

)

for 0,5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

< 5 % U

T

(>95 % dip in U

T

)

for 5 sec

Mains power quality should be
that of a
typical commercial or hospital
environment. If the user of the
PENTA requires
continued operation during
power
mains interruptions, it is
recommended
that the PENTA be powered from
an uninterruptible power supply
or a battery.