Port RX (Electrical Interfaces)

Frequency Analysis

The FTB-8100 Series allows the following frequency monitoring range.

³

Actual Frequency (bps): Indicates the frequency of the input signal.

³

Frequency Offset: Indicates the offset between the standard rate
specification and the rate of the input signal.

³

Max. Positive Offset: Indicates the offset between the standard rate
specification and the largest rate recorded from the received signal.

³

Max. Negative Offset: Indicates the offset between the standard rate
specification and the smallest rate recorded from the received signal.

Offset Unit: Allows the selection of the frequency offset unit. Choices are
bps and ppm. The default setting is ppm.

Interface

Standard Rate Specification

DS1

1544000 ±140 ppm

E1

2048000 ±100 ppm

E2

8448000 ±100 ppm

E3

34368000 ±100 ppm

DS3

44736000 ±100 ppm

STS-1e/STM-0e

51840000 ±100 ppm

E4

139264000 ±100 ppm

STS-3e/STM-1e

155520000 ±100 ppm