Quick start guide, How to perform the first titration – Hanna Instruments HI 903 User Manual

Page 11

Advertising
background image

11

QUICK START GUIDE

HOW TO PERFORM THE FIRST TITRATION

Method Selection

For this analysis we will use the HI8301EN Solvent with 5mg/ml 1-component
Titrant
standard method

.

To select this method:

• Press

from the

Idle

screen. Use the

and

keys to highlight the

HI8301EN Solvent with 5mg/ml 1-component Titrant method.

• Press

.

After accomplishing these operations, the method’s name will be displayed on the top line of
the

Idle

screen.

Setting Method Parameters

To display the method parameters, press

. The

View/Modify Method

screen will be

displayed.
Only certain parameters from the standard methods can be changed.
For this titration, only the KF titrant concentration value and the size of the solvent sample
need to be entered as in the screen shown below.

To accomplish this:

• Highlight

Titrant

option from the

View/Modify Method

screen and then press

.

• The

Karl Fischer Titrants

screen will be displayed.

• Press

.

• Highlight

“Standardized Titrant Concentration”

and

press

.

• Input the correct value, then press

.

• Press

three times to return to the

Idle

screen.

Setup Titration Report

Users can select the information that is stored for each titration that is performed.
To obtain proper information at the end of the titration, perform the following operations:

• From the main screen, press

and the

Data Parameters

screen will be

displayed.