Hi8001en_r0.4, Method id: hi8001en – Hanna Instruments HI 903 User Manual

Page 163

Advertising
background image

Method ID: HI8001EN

5.0 mg/mL Titrant Standardization using a Liquid Water Standard

for one-component titrant

Titrator Application Methods: General Methods - r. 0.4

Description:

Method for the standardization of 5.0 mg/mL One-
Component Karl Fischer Titrant using a Liquid Water
Standard. The results are expressed in mg/mL.

Electrode:

HI 76320 Double Platinum Pin Electrode


Reagents:

5 mg/mL One-Component KF Volumetric Titrant

10 mg/g Liquid Water Standard

Dry Methanol


Other Accessories:

A clean, dry 3-mL syringe

A clean, dry 22-gauge

, 6” non-coring septum

penetration needle

GL 45-Thread Solvent Bottle

Procedure:

Setup titrator according to the instruction manual.

Press “Select Method” from the main screen. Use
the arrow keys to highlight ‘HI8001EN Stdz
5mg/mL w/ water Std.

’ and press “Select”.

Install a 5-mL burette filled with 5 mg/mL One-

Component

KF

Volumetric

Titrant

to

be

standardized and verify that no air bubbles are
present in the burette or tubing. If necessary prime
until all air has been removed completely.

Connect the solvent bottle top assembly to the

bottle of methanol according to the manual.

Prepare the titration vessel according to the

manual. Dispense enough solvent from the solvent
bottle to fill the vessel to the “min” line (about 50
mL).

Press “Start/Stop” to pre-titrate the solvent and the
titration vessel moisture. Allow the background
drift rate to stabilize before proceeding to the next
step.

Fill the syringe and needle with the water

standard.

Weigh the syringe, needle and water standard.

Press “Start Analysis”. You will be prompted to
enter the sample size.

Dispense 1.00 g (about 1 mL) of standard into the

titration vessel through the septum using the
needle. Pay attention not to get any sample on
the electrode or beaker walls. If necessary swirl
the titration vessel by hand.

Clear the needle of residual standard by intaking a

small volume of air from the titration vessel. If a
“hanging drop” of standard is seen on the end of
the needle, dip the end of the needle briefly in the
solvent.

Remove the needle from the titration vessel and

weigh the syringe again in order to determine the

added standard mass (by difference of the two
measurements.)

Use the numeric keypad to enter the exact weight

and press “Enter” to start the analysis.

At the end of the titration,

the “Standardization

Result

” screen is displayed. The results are

expressed in mg/mL.


Method Parameters:

Name:

Stdz 5mg/mL w/ water Std

Method Revision:

1.0

Type:

Titrant Standardization

Predispensing Amount:

25 %

Pre-Analysis Stir Time:

5 Sec

Stirring Speed:

900 RPM

Stirbar Type:

Medium

Drift Entry:

Automatic

Solvent:

Methanol

Standard:

Liquid 10 mg/g

Type:

Liquid by mass

Concentration Unit:

mg/g

Water Content:

10.0000 mg/g

Standard Size:

1.0000 g

Titrant:

Composite 5

Titrant Type:

One Component

Nominal Titrant Conc.:

5.0000 mg/mL

Std. Titrant Conc.:

5.0000 mg/mL

Date/Time:

Jan 1, 2011 12:00

Titrant Age Reminder:

2d:00h:00m

Control Parameters:

Start Mode:

Normal

Standby Mode:

Enabled

Standby Duration:

720 minutes

Imposed Current:

20 uA

Minimum Dose:

0.500 uL

Maximum Dose:

20.000 uL

Timed Increment:

1 second

End Point Value:

180.0 mV

Signal Averaging:

3 Readings

Flow Rate:

10.0 mL/min

Termination Parameters:

Maximum Duration:

1200 sec

Maximum Titrant Volume:

10.0000 mL

Term. Criterion:

Relative Drift

Relative Drift:

7.0 ug/min

Calculations:

Titrant units:

mg/mL

Titrant volume consumed:

V (mL)

Final Results Units:

mg/mL

Standard Concentration:

10.0000 mg/g

Standard mass:

1.0000 g