5.2
EYE-C-GAS Operation
5.2.1 General
5.2.1.1 Power On & Off
Before powering on the camera, ensure that the battery is properly inserted and fully charged.
EYE-C-GAS Power Button: (see Figure 18 below)
To power on the EYE-C-GAS camera, depress the middle Power/Menu Button for one second. A
sequence of screens will appear as the camera initiates an internal calibration sequence.
During the initial 30 second (approximate) power up sequence, EYE-C-GAS will display a white
screen followed by a color test screen. After this is completed, the Opgal logo will appear and
will fade to black (the battery LEDs will also fade) during the next 30 seconds to conserve battery
life while the internal cooler lowers to its required operating temperature. After the LCD dims to
black, the leftmost LED battery indicator will blink indicating that the detector cooler is cooling
to its required temperature.
The cool down sequence will take approximately 8 to 10 minutes. When completed, a video
image will appear on the LCD. To power off the EYE-C-GAS, depress the Power/Menu Button
for a minimum of 8 seconds.
Blinking LED
Figure 19: Logo and LED blinking during cool down
Figure 18: Keypad, multifunction middle button
Middle Button
Power On & Off.
Also Menu and
Select