From the Normal mode when the Mode button is pressed the keypad

configuration will change to provide adjustments in the Enhanced Mode as
shown in Figure 22. The adjustment is required for maximum detection while
having a clear and steady image. The default values provide full detection
capabilities, but in order to bring the camera to a clear and steady image without
losing the gas leak detection change to lower values Enh+/Enh- first and then
Diff+/Diff- as shown in Table 1 on page 24.

Figure 22: Enhanced Mode of operation

Figure 21: Keypad configuration in Enhanced Mode operation