Caution – Heath Consultants Remote Methane Leak Detector (RML) User Manual

Page 22

Advertising
background image

When the Self-Test fails. The following table may help to isolate the problem:

Cause

Corrective Action

Low signal strength or noisy

Make sure battery is fully charged

Make sure that the unit is in its proper position, and

all the way down and flat

Initial ppm-m too low or too
high

Make sure battery is fully charged

Make sure the unit is in its proper position, all the

way down, and flat

Verify calibration cell is not broken

Use the Calibration Override

Laser adjustment failed

Make sure battery is fully charged

Make sure that the unit is in its proper position, and

all the way down and flat

Verify calibration cell is not broken

**Caution**

Should the instrument not pass after several attempts, do not use the instrument for
survey work until the problem is corrected. Contact HEATH for further assistance.

Page 22 of 36

101515-0 RMLD MANUAL REV E