Chapter v, Maintenance and troubleshooting information – Heath Consultants Remote Methane Leak Detector (RML) User Manual

Page 29

Advertising
background image

Chapter V

Maintenance and Troubleshooting Information

Troubleshooting the Instrument

The advanced design of the RMLD makes it one of the most reliable leak survey
instruments available. However, should you experience problems with the instrument or
suspect that the instrument is not operating properly, do not use the instrument for leak
survey work until the problem is resolved.

Only a qualified RMLD repair technician should attempt to repair or adjust the
instrument. There are no user serviceable components in the RMLD that can be repaired
or replaced.

Most often, a problem has a simple cause. The following table provides a list of common
problems, cause and solution. Should you have a problem not listed or the solution
doesn’t work, please contact HEATH for further assistance.

Symptom

Probable Cause(s)

Solution

Readings are higher than
normal at short range and
lower than normal for long
range.

• Laser calibration has

drifted

• Run Self-Test

Concentration reading low
and will not pass the Self-
Test

• Laser calibration has

drifted

• Run Self-Test and use

the override function

Have to use Self-Test
override each time the Self-
Test is performed

• Laser calibration has

drifted

• Laser calibration drift

rate is too high

• Perform the Self-Test

on a daily basis

• Contact HEATH.

Unit will not turn on

• Low battery

• Recharge internal

battery, or

• Use the External Power

Supply

Keep getting a warning
sound or icon when
scanning

• Scanning at a distance

beyond the RMLD
range

• Background surface is

absorbing or reflection
the light level

• Move closer to the

target, or

• Change angle to the

target to get a better
reflecting background

• Check battery level and

Page 29 of 36

101515-0 RMLD MANUAL REV E

Advertising