Heath Consultants Remote Methane Leak Detector (RML) User Manual

Page 30

Advertising
background image

• Low battery

recharge if necessary

Excessive false detections
while scanning at longer
distances

• Scanning too fast

• Alarm detection

threshold set too low

• Scanning at the range

limit of the instrument

• Slow down the scanning

rate. Pause at the long
range and sweep

towards you

• Increase the Alarm

detection threshold

• Move in closer

Excessive false detections
while scanning at closer
distances

• Scanning too fast

• Alarm detection

threshold set too low

• Avoid making abrupt

motions while scanning

• Increase the Alarm

Detection Threshold

Excessive false detections
or loss of sensitivity

• Laser output not

optimized

• Alarm Detection

Threshold set too high
or low for conditions

• Perform Self-Test

procedure

• Check the Alarm

Detection Threshold

Error icon or Warning icon
on continuously

• Low battery

• Moisture condensation

on mirror due to rapid
change in temperature

• Internal component

failure

• Check battery level and

recharge if necessary

• Allow for the

temperature of the
instrument to stabilize

• Note error code and

contact HEATH

Battery indicator does not
show full charge after
charging

• Battery not fully

charged

• Battery level calibration

error

• Charge unit until solid

green light on charger is
on

• Run unit until it shuts

off, then fully recharge
with out interruption

Battery indicator bars are
flashing on and off

• Battery level calibration

error

• Run unit until it shuts

off, then fully recharge
with out interruption

Page 30 of 36

101515-0 RMLD MANUAL REV E