Use of the Clean Bench


Planning

• Thoroughly understand procedures and equipment before beginning work.

• Arrange for minimal disruptions, such as room traffic or entry into the

room, while the clean bench is in use.

Start-Up

• Turn on clean bench blower.

• Check the prefilter for obstructions.

• Wipe down the interior surfaces of the clean bench with stainless steel

cleaner or 70% ethanol. DO NOT use abrasive cleaners, bleach or
solvents, as they may damage the work surface of the clean bench.

• Allow the clean bench to operate undisturbed for 5 to 15 minutes before

loading materials.

• Wear long sleeved lab coat with knit cuffs and over-the-cuff rubber

gloves. Use protective eyewear.


Loading Materials and Equipment

• Only load the materials required for the procedure. Do not overload the

clean bench.

• Do not obstruct the air diffuser.

• Large objects should not be placed close together.

• After loading the clean bench, wait 2 to 3 minutes to purge airborne

contaminants from the work area.

Work Techniques

• Keep all materials at least 4 inches inside of the clean bench and perform

all contaminated operations as far to the front of the work area as possible.

• Perform sterile operations as far to the rear of the work area as possible.

• Segregate all clean and contaminated materials in the work area.

• Arrange materials to minimize the movement of contaminated materials

into clean areas.

• Keep all discarded contaminated material to the front of the clean bench.

• Avoid moving materials or the operator’s hands and arms in and out of the

work area during use.

• Avoid the use of an open flame.

• Use proper aseptic technique. – “Sterile Air First”

• Avoid using techniques or procedures that disrupt the airflow patterns of

the clean bench.

Final Purging

• Upon completion of work, the clean bench should be allowed to operate

for 2 to 3 minutes undisturbed, to purge airborne contaminants from the
work area.