Refer to the following table if your Purifier XT Clean Bench fails to operate
properly. If the suggested corrective actions do not solve your problem, contact
Labconco Product Service for additional assistance.

PROBLEM

CAUSE

CORRECTIVE ACTION

Blower and lights
won’t turn on

Unit not plugged into
outlet

Plug the clean bench into appropriate
electrical service.

Circuit breakers tripped

Reset circuit breakers.

Motor won’t run

Speed control out of
adjustment

Have speed control adjusted by a
qualified certifier.

Defective speed control

Replace speed control.

Contamination of
work in the clean
bench

Improper technique or
procedure for the clean
bench

See “Use of the XT Clean Bench”
section of this manual.

Contamination of
work in the clean
bench (cont.)

External factors are
disrupting the clean
bench airflow patterns
or acting as a source of
contamination

See “Installation” section of this
manual.

Clean bench is out of
adjustment/HEPA
filter(s) are defective

Have clean bench recertified.

Excessive vibration

Motor/blower out of
adjustment

Repair/Replace Motor/Blower Assy.

Application requires
equipment isolation

Accessory vibration table.