Chapter 6: Maintaining Your XT Clean Bench

Product Service 1-800-522-7658

29

Routine Maintenance Schedule


Under normal operation, your Purifier XT Clean Bench will require little routine
maintenance. The following schedule is recommended:

Weekly

• Wipe down the interior surfaces of the clean bench with stainless steel

cleaner or 70% ethanol and allow to dry.

• Using a damp cloth, clean the exterior surfaces of the clean bench,

particularly the front and top of the clean bench to remove any
accumulated dust.


Monthly (or more often as required)

• Check all service valves, if so equipped, for proper operation.

• Check the prefilters and replace if necessary. The prefilters should be

replaced at least quarterly.

• All weekly activities.


Quarterly

• Replace the prefilter. See Appendix A: XT Clean Bench Components for

ordering information.

• All monthly activities.


Annually

• Have the clean bench recertified by a qualified certification technician.

• All quarterly activities.