Chapter 6: Maintaining Your XT Clean Bench

Product Service 1-800-522-7658

34

Motor/Blower Replacement:

A QUALIFIED CERTIFICATION TECHNICIAN SHOULD
SERVICE THE MOTOR/BLOWER. FOLLOWING
REPLACEMENT OF A MOTOR/BLOWER, A QUALIFIED
CERTIFICATION TECHNICIAN MUST RECERTIFY THE
CLEAN BENCH.

Do NOT contact blower wheel while still in motion.
NE PAS être en contact avec la roué du ventilateur tant qu’il est
en marche.


1. Unplug the clean bench.
2. Remove the prefilters, as described earlier.
3. Remove the side windows, and worksurface, as described in Chapter 3,

Section: Moving XT Clean Bench through Tight Doorway

4. Remove (3) Torx Head Screws to remove Blower Compartment Panel, as

shown in Figure 6-5. NOTE!!! Two other screws will be left in place.

5. Disconnect the two wiring harnesses from the head of the motor.
6. Remove two bolts on the blower housing flange (left side).

The motor/blower assembly is heavy. Lift & handle with care.
L’ensemble moteur/ventilateur est lourd. Soulever et manipuler
avec precaution.

7. Pull Motor Blower Assembly at an angle to remove right side flange from

inside plenum compartment.

8. Remove Motor Blower Assembly from Blower Compartment.
9. To replace the motor/blower, reverse the above steps.
10. Recertify the unit before use.

)

!

!

Figure 6-5

Blower

Compartment

Panel

(3) Torx Head

Screws from here

Blower Motor

Assembly