Martin Christ Epsilon 2-6D LSCplus User Manual

Page 106

Advertising
background image

Freeze-dryer EPSILON 2-6 LSCplus

10 Technical data

106

Version 10/2012, Rev. 1.6 of 16/12/2014 • sb

Translation of the original operating manual

Connection requirements

Electrical connection:

3 x 400 V / 50 Hz
(other supply data on request)

Protection class:

I

IP-Code according to DIN 60529:

21

Nominal power:

3.45 kVA

Nominal current:

8.0 A

Mains fuse:

12 A

Equipment connections

Vacuum connection:

Small flange connection DN25KF
(ISO 28403, DIN 2861)

Media drain valve:

Hose nozzle DN10 (outside diameter 12 mm)

Aeration valve:

Hose nozzle DN6 (outside diameter 8.5 mm max)

Mains input:

Cekon CEE (16 A 3ph)

Data interface (LAN):

RJ 45

Special equipment: water cooling system

Part number:

113120

Water feed flow temperature:

+15°C to +20°C

Water return flow temperature:

25°C max.

Water temperature difference:

10K max. with +15°C feed flow temperature

5K max. with +20°C feed flow temperature

Cooling water pressure:

10 bar max.

Differential pressure between feed flow and return
flow:


1.5 bar max.

Volumetric flow rate:

0.38 m

3

/h max.

Heat dissipated by cooling water:

3.065 kW

Cooling water feed flow connection:

R3/4" with hose nozzle DN 13

Cooling water return flow connection:

R3/4" with hose nozzle DN 13

Option: cooling water flow controller (factory setting):

12 bar (refrigeration unit, high pressure side)