Freeze-dryer EPSILON 2-6 LSCplus

6 Operation

64

Version 10/2012, Rev. 1.6 of 16/12/2014 • sb

Translation of the original operating manual

Failure memory (2)
The failure memory stores the most recent messages of the process and
equipment information system. These messages can be viewed in this
dialog box. The failure memory includes the last 32 messages. If this
number is exceeded, the oldest message will be overwritten.
Use the arrow keys to call up the individual messages.
The error message text is always followed by an error code.

NOTE

Always indicate the error code in the event of enquiries or service
requests!

Fig. 55: Dialog box "Failure memory"


Vacuum pump oil change (4)
The system monitors the oil change interval of the vacuum pump. The
interval can be adapted to the vacuum pump model and utilisation.
When the end of an oil change interval is reached, a corresponding
message will be displayed.
• Acknowledge the message.
• Change the oil of the vacuum pump.
• Reset the operating hour counter in the dialog box “Vacuum pump oil

change” by pressing the “reset” button.