Freeze-dryer EPSILON 2-6 LSCplus

5 Set-up and connection

Version 10/2012, Rev. 1.6 of 16/12/2014 • sb

35

Translation of the original operating manual

Pos: 100 /20 0 Ch rist/3 61 G T-BA L abo r-Pilot (PROJEKTE)/Epsilo n Sta nda rd/0 50 Aufs tellun g u nd Ansc hluss/0 50- 003 0 Tro ckne rpat ron e (Ad sor ptionsfilt er ) - Ü bersc hrift -- --- -- --- --- -- --- -- --- -- --- - @ 26\m od_ 140 505 799 428 5_68 .docx @ 1 909 60 @ 2 @ 1

5.3 Drying cartridge (adsorption filter)

Pos: 101 /20 0 Ch rist/3 61 G T-BA L abo r-Pilot (PROJEKTE)/Epsilo n Sta nda rd/0 50 Aufs tellun g u nd Ansc hluss/0 50- 003 2 Tro ckne rpat ron e Epsilo n (n ur Ve ntil) - T ext @ 25\ mod _14 049 846 256 67_6 8.d ocx @ 185 693 @ @ 1

The drying cartridge prevents the air humidity from being absorbed by the
heat transfer medium.
During the transport of the freeze-dryer, the drying cartridge is removed.
Prior to commissioning the unit, it must be installed.
• Remove the sealing cap (1) from the valve.
• Connect the supplied adapter (5) together with the drying cartridge (4)

to the valve (2).

• Open the valve (2) by turning it anti-clockwise up to the stop.


1 Sealing cap
2 Valve
3 Connection of the

expansion vessel

(inside the unit)

Fig. 16: Valve with sealing cap for transport


4 Drying cartridge

(adsorption filter)

5 Adapter

Fig. 17: Valve with drying cartridge (adsorption filter)

NOTE

The silica gel of the drying cartridge has a limited adsorption capacity.
If the colour of the silica gel changes from orange to green (see label on
the drying cartridge), the drying cartridge must be replaced (Christ order
no. 610 990).

Pos: 102 /01 0 Univ ersal mod ule/Seit enwec hsel @ 0\m od_ 120 211 624 4312 _0. docx @ 10 5 @ @ 1