Freeze-dryer EPSILON 2-6 LSCplus

6 Operation

Version 10/2012, Rev. 1.6 of 16/12/2014 • sb

75

Translation of the original operating manual

• Pressing the button “Program name” (2) calls up an input field in which

the name of the program can be changed.

• The button “Save” (7) can be used in between or at the end of the

program creation in order to save the program.

• The button “Program” (1) calls up the main window. The system will ask

the user whether the program shall be saved if this has not been done
yet.

The program will be automatically assigned to the first free program storage
location. The creation of the program is now complete and it can be loaded.

Button ">>more"
The button ">>more" can be used to enter more functions and set values in
a program-related manner.

Fig. 68: Selection of program-related functions and set values


Vacuum pump warm-up
see chapter 6.5.3 - "Freeze-drying with the PGMplus programmer module"

Option: Pressure increase test (only with the double-chamber method)
The pressure increase test can only be performed when the freeze-dryer
is equipped with an intermediate valve.

NOTE

The indication of the measurement value "dp Test" can be configured in
chapter 6.5.1.1 - "Main window "Manual"" under "Actual values of the
current process" in the dialog box "Select actual value".


In contrast to the pressure increase test in the manual mode, repeated
pressing of the button in the program mode enables the selection of
different variants.