6 Operation

Version 04/2012, Rev. 1.3 of 25/11/2013 • sb

31

Translation of the original operating manual

Pos: 114 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /06 0 Betri eb/0 60- 004 0-0 030 Einsetz en v on Zub ehö r @ 8 \mo d_1 309 508 120 119 _68. docx @ 46 850 @ 3 @ 1

6.4.3 Installation of accessories

• Only use inserts that are suitable for the rotor (see chapter 11.1 -

"Rotor program").

• In swing-out rotors, all places of a rotor must be loaded with buckets..
• Always load the opposite inserts of the rotors with the same accessories

and fill to avoid imbalance.


Evaporation with different tube sizes
Working with different tube sizes is possible. In this case, however, it is
very important that the inserts are installed symmetrically (see figure).

Fig. 10: Permissible loading of the swing-out rotor with different tube sizes


Evaporation with low capacity
The tubes must be installed symmetrically so that the buckets and their
inserts are loaded evenly to avoid imbalance.

correct

incorrect