Martin Christ RVC 2-18 CDplus User Manual

Page 74

Advertising
background image

RVC 2-18 CD&

12 Glossary

74

Version 04/2012, Rev. 1.3 of 25/11/2013 • sb

Translation of the original operating manual

Safety pressure
Since the vacuum has a dominating influence on the product temperature,
Martin Christ Gefriertrocknungsanlagen GmbH has integrated a so-called
safety pressure function into their units in order to protect the products. If
the pressure inside the rotor chamber rises too quickly – above the safety
limit – the energy supply of the unit will be interrupted so that the
evaporation process slows down.

Fig. 48: Diagram describing the safety pressure function

=== E nde der Liste für Tex tma rke I nhal t = ==

Advertising