8 Maintenance and service

54

Version 04/2012, Rev. 1.3 of 25/11/2013 • sb

Translation of the original operating manual

• Push the drive shaft onto the axle.
• Install the thrust piece and the compression spring on the axle and

secure them in place with the countersunk screw with 1.5 Nm.

• Push the cap fully onto the drive shaft.
• Connect the power cord and switch the RVC on.
• Perform a test run.

NOTE

If the unit does not reach the necessary speed, the gap between the
magnetic coupling and rotor chamber bottom may be too wide (optimum
value: 2.4 0.3mm).
• In this case, press the rotor shaft down!

Pos: 160 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1

Pos: 161 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1

Pos: 162 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /08 0 Wart ung un d Inst and haltu ng/ 080 -001 0- 0030 Z ube hör @ 8\ mo d_1 3100 265 340 81_ 68.d ocx @ 472 71 @ 3 @ 1

8.1.3 Accessories

CAUTION

For the care of the accessories, special safety measures must be
considered as these are measures that will ensure operational safety at
the same time.


Chemical reactions as well as stress-corrosion (combination of oscillating
pressure and chemical reaction) can affect or destroy the metals. Barely
detectable cracks on the surface can expand and weaken the material
without any visible signs.
• Check the material regularly (at least once a month) for

– cracks
– visible damage of the surface
– pressure marks
– signs of corrosion
– other changes.

• Replace any damaged components immediately for your own safety.
• Immediately rinse off the rotors or accessories if any liquids that may

cause corrosion come into contact with them.

• Clean the accessories outside the rotational vacuum concentrator once

a week or preferably after each use.

• If the rotors or accessories have been contaminated with toxic,

radioactive, or pathogenic substances, clean them immediately with a
suitable decontamination agent (depending on the type of
contamination). Take suitable precautions for your own safety if there is
a risk of toxic, radioactive, or pathogenic contamination.