Switching functions

SP1, SP2

Setting range

0 … 99% F.S. (0 … 9.9 V)

Hysteresis

1% F.S.

Relay contact

30 VDC/

≤0.5 ADC

floating (n.o.)

closed

p

≤ p

SP

(LED lit solid)

open

p

≥ p

SP

(LED off)

Switching time

≤50 ms

Status relay

closed

measurement mode
warning

open

no supply voltage
warming up
error

RS232C interface

Transmission rate
Data format


9600 baud
binary
8 data bits
one stop bit
no parity bit
no handshake
→ "Power Connection"

For further information on the RS232C interface

 [1].

Supply

DANGER

The gauge may only be connected to power sup-
plies, instruments or control devices that conform
to the requirements of a grounded protective extra-
low voltage (SELV) and limited power source
(LPS), Class 2. The connection to the gauge has to
be fused.