3 switching functions – Nor-Cal CDG 100 OP Lit User Manual

Page 27

Advertising
background image

CDG100-OP-LIT (2010-1)

27

Push the <ZERO> button again:

Fine adjustment
within 0...3 s:

the zero adjustment value changes
by one unit (push <ZERO> button in
intervals of 1 s)

Change of direction
within 3...5 s:

the zero adjustment changes its
direction (the blinking frequency of
the <STATUS> LED changes
briefly)

If the <ZERO> button is released for more than 5 s, the

gauge returns to the measurement mode.

The <STATUS> LED blinks green if the signal output is negative
(< -20 mV).

4.3

Switching Functions

The two switching functions can be set to any pressure within
the measurement range of the gauge (

 14).

The current setpoint setting
• is output at the D-Sub connector instead of the measurement

signal (

 20) and can be measured with a voltmeter after

the <SP> button is pressed, or

• can be read/written via the RS232C interface.

If the pressure is lower than the setpoint, the corresponding LED
(<SP1> or <SP2>) is lit solid and the corresponding relay
(

 20) is energized.