2 (zero( adjustment with ramp function – Nor-Cal CDG 100 OP Lit User Manual

Page 25

Advertising
background image

CDG100-OP-LIT (2010-1)

25

Briefly press the <ZERO> button with a pin (max.

ø1.1 mm). The zero adjustment runs automatically. The
<STATUS> LED blinks until the adjustment (duration

≤8 s)

is completed.

Press the button briefly

max. ø1.1 mm

After zero adjustment, the gauge automatically returns

to the measurement mode.

The <STATUS> LED blinks green if

• the signal output is negative (< -20 mV) when the final pres-

sure has been attained

• the zero adjustment has failed.

4.2.2

<ZERO> Adjustment with Ramp Function

The ramp function allows to adjust the zero at a known reference
pressure within the measurement range of the gauge.

It also permits to adjust an offset of the characteristic curve in
order to
• compensate for the offset of the measuring system or
• obtain a slightly positive zero for a 0 … 10 V AD converter.