Nor-Cal CDG 100 OP Lit User Manual

Page 26

Advertising
background image

26

CDG100-OP-LIT (2010-1)

The offset should not exceed 2% of the F.S. (+200 mV). At a
higher positive offset, the upper limit of the measurement range
is exceeded.

Zero adjustment using the ramp function can be

performed via
• the <ZERO> button on the gauge,
• the RS232C interface (→  [1]).

Recommended procedure for adjusting the offset of a

measuring system:

→ Notice  23.

Operate the gauge for at least 90 minutes under constant

ambient conditions (until the signal is stable).

Push the <ZERO> button with a pin (max. ø1.1 mm) and

keep it depressed. The <STATUS> LED starts blinking.
After 5 s, the zero adjustment value, starting at the current
output value, keeps continually changing (ramp) until the
button is released or until the setting limit (max. 25% F.S.)
is reached. The corresponding output signal is delayed by
about 1 s.

Keep the button depressed

max. ø1.1 mm

Advertising