4.2.1

<ZERO> Adjustment

The zero can be adjusted via

• the <ZERO> button on the gauge,
• the digital input "Remote Zero" (briefly apply the

supply voltage (+14 … +30 V) to pin 10),

• the RS232C interface (→  [1]),
• a Nor-Cal Vacuum Gauge Controller.

While the gauge is being heated and/or under atmos-

pheric pressure, the zeroing function is locked in order
for operating errors to be prevented.

Evacuate the gauge to a pressure according to the table

below:

F.S.

Recommended final pressure
for zero adjustment

1000 Torr
100 Torr

20 Torr

10 Torr

2 Torr

1 Torr

0.1 Torr

<5

×10

-2

Torr

<5

×10

-3

Torr

<1

×10

-3

Torr

<5

×10

-4

Torr

<1

×10

-4

Torr

<5

×10

-5

Torr

<5

×10

-6

Torr

If the final pressure is too high for zero adjustment (>25%
of the F.S.), the zero cannot be reached and the
<STATUS> LED blinks green. If this is the case, activate
the factory setting and adjust the zero again (

 30).

Operate the gauge for at least 90 minutes under constant

ambient conditions (until the signal is stable).