Good F2 fuse: 00.0

Ω

to

00.5

Ω

Good F1 fuse: 0.995 k

Ω

to

1.005 k

Ω

Replace fuse: OL

Replace fuse: OL

Touch top half

of input contacts

aom5f.eps

Figure 10. Testing the Current Fuses

Replacing the Battery

Replace the battery with a 9 V battery (NEDA A1604,
6F22, or 006P).

X

W

Warning

To avoid false readings, which could lead to
possible electric shock or personal injury,
replace the battery as soon as the battery
indicator (
b) appears. If the display shows

bAtt the Meter will not function until the

battery is replaced.

Replace the battery as follows, refer to Figure 11:
1. Turn the rotary switch to OFF and remove the test

leads from the terminals.

2. Remove the battery door by using a standard-blade

screwdriver to turn the battery door screws one-
quarter turn counterclockwise.

3. Replace the battery and the battery door. Secure the

door by turning the screws one-quarter turn
clockwise.