Using measure mode, Measuring electrical parameters (upper display), Current measurement with loop power – Fluke 725 User Manual

Page 27: Figure 7. measuring voltage and current output

Advertising
background image

Using Measure Mode

17

Using Measure Mode

Measuring Electrical Parameters (Upper Display)

To measure the current or voltage output of a transmitter,
or to measure the output of a pressure instrument, use the
upper display and proceed as follows:

1. Press

l

to select volts or current. LOOP should

not be on.

2. Connect the leads as shown in Figure 7.

Current Measurement with Loop Power

The loop power function activates a 24 V supply in series
with the current measuring circuit, allowing you to test a
transmitter when it is disconnected from plant wiring. To
measure current with loop power, proceed as follows:

1. Connect the calibrator to the transmitter current loop

terminals as shown in Figure 8.

2. Press

l

while the calibrator is in current

measurement mode. LOOP appears and an internal
24 V loop supply turns on.

V mA

LOOP

V mA

TC

RTD

˚

C

˚

F

Hz

100%

25%

25%

RECALL

ZERO

MEAS

SOURCE

STORE

SETUP

0%

MULTIFUNCTION CALIBRATOR

725

Black

Red

sh42f.eps

Figure 7. Measuring Voltage and Current Output