Testing an output device, For current or dc voltage, or, For source mode – Fluke 725 User Manual

Page 55: Figure 22. calibrating a chart recorder

Advertising
background image

Testing an Output Device

45

Testing an Output Device

Use the source functions to test and calibrate actuators,
recording, and indicating devices. Proceed as follows:

1. Connect the test leads to the instrument under test

as shown in Figure 22.

2. Press

V

for current or dc voltage, or

F

for

frequency or resistance (lower display).

3. If necessary, press

M

for SOURCE mode.

V mA

LOOP

V mA

TC

RTD

˚

C

˚

F

Hz

100%

25%

25%

RECALL

ZERO

MEAS

SOURCE

STORE

SETUP

0%

MULTIFUNCTION CALIBRATOR

725

Black

Red

0 to 1 V dc

Input

sh25f.eps

Figure 22. Calibrating a Chart Recorder