True-rms Digital Multimeters

Detailed Specifications

69

AC Current Specifications

Accuracy

Function Range Resolution

20 to 45 Hz

45 to 1 kHz

1 to 20 kHz

20 to 100 kHz

[4]

500

μ

A 0.01

μ

A

1 % + 20

0.6 % + 20

0.6 % + 20

5 % + 40

AC

μ

A

[3]

5000

μ

A 0.1

μ

A

1 % + 5

0.6 % + 5

0.6 % + 10

5 % + 40

50 mA

0.001 mA

1 % + 20

0.6 % + 20

0.6 % + 20

5 % + 40

AC mA

[3]

400 mA

0.01 mA

1 % + 5

0.6 % + 5

1.5 % + 10

5 % + 40

5 A

0.0001 A

1.5 % + 20

0.8 % + 20

3 % + 40

[4]

Not

Spec’d

AC A

[2]

10 A

[1]

0.001 A

1.5 % + 5

0.8 % + 5

3 % + 10

[4]

Not

Spec’d

[1]

10 A range (10 % to 100 % of range).

[2]

20 A for 30 seconds on, 10 minutes off. >10 A not specified.

[3]

400 mA continuous; 550 mA for 2 minutes on, 1 minute off.

[4]

Verified by design and type tests.

See Detailed Specifications introduction for additional information.