Frequency counter sensitivity – Fluke 289 User Manual

Page 85

Advertising
background image

True-rms Digital Multimeters

Detailed Specifications

75

Frequency Counter Sensitivity

Approximate Voltage Sensitivity

(rms sine wave)

[1]

Input Range

15 Hz to 100 kHz

AC Bandwidth

[2]

Approximate DC

Trigger Levels

DC Bandwidth

[2]

50 mV

5 mV

1 MHz

5 mV & 20 mV

600 kHz

500 mV

25 mV

1 MHz

20 mV & 60 mV

1 MHz

5 V

0.25 V

700 kHz

1.4 V & 2.0 V

80 kHz

50 V

2.5 V

1 MHz

0.5 V & 6.5 V

1 MHz

500 V

25 V

300 kHz

5 V & 40 V

300 kHz

1000 V

50 V

300 kHz

5 V & 100 V

300 kHz

Approximate Current Sensitivity

(rms sine wave)

Input Range

15 Hz to 10 kHz

AC Bandwidth

Approximate DC Trigger

Levels

DC Bandwidth

500

μ

A 25

μ

A

100 kHz

5000

μ

A 250

μ

A

100 kHz

50 mA

2.5 mA

100 kHz

400 mA

25 mA

100 kHz

5 A

0.25 A

100 kHz

10 A

1.0 A

100 kHz

NA NA

[1]

Maximum input = 10 x range (1000 V maximum, 2 x 10

7

V-Hz product maximum). Noise at low frequencies and amplitudes may affect accuracy.

[2]

Typical frequency bandwidth with full scale (or maximum 2 x 10

7

V-Hz product) rms sine wave.

Advertising
This manual is related to the following products:

287