English – Abbott Freestyle Freedom Lite User Manual

Page 21

Advertising
background image

17

Message

What It Means

What To Do

There may be a problem

with the test strip

Meter error

Very high blood glucose

higher than 500 mg/dL

HIGH control solution

was applied when the

temperature was too

cold . (Applies only to

control solution labeled

as HIGH) .

1 . Follow your healthcare professional’s advice to

treat high blood glucose if you have symptoms

such as:

Fatigue

Thirst

Excess urination

Blurry vision

2 . Perform a control solution test using a new

test strip . If the test results are within the range

printed on the test strip vial, retest using blood

and a new test strip .

3 . If the control solution result is out of range or the

error reappears, contact Customer Care .

Contact Customer Care Toll Free at 1-888-522-5226 www.MyFreeStyle.com