Message

What It Means

What To Do

Incorrect test procedure .

For example, putting

blood on the test strip

before inserting it into

the meter or applying
blood before

and

appear

There may be a problem

with the test strip

Meter error

1 . Wait for

and to appear before applying

blood or control solution .

2 . Perform a control solution test using a new

test strip . If the test results are within the range

printed on the test strip vial, retest using blood

and a new test strip .

3 . If the control solution result is out of range or the

error reappears, contact Customer Care .

There may be a problem

with the test strip

Meter error

1 . Perform a control solution test using a new

test strip . If the test results are within the range

printed on the test strip vial, retest using blood

and a new test strip .

2 . If the control solution result is out of range or the

error reappears, contact Customer Care .