English – Abbott Freestyle Freedom Lite User Manual

Page 28

Advertising
background image

24

4.

Confirm System Check Screen .

This display appears when you turn on the meter . Do not use the meter if

the display check screen does not exactly match the example. Contact

Customer Care.

See The FreeStyle Freedom Lite® Meter and How It Works section for more

information .

If the system check screen does not appear, go to Troubleshooting section .

5.

Apply and hold control solution to 1 sample area of the test strip until:

You see short lines moving clockwise on the display or

You hear a beep

This indicates the test strip obtained enough control solution and the meter is

processing . If the test does not start after applying the control solution, go to

Troubleshooting section .

IMPORTANT: If a short line does not appear after 5 seconds, the

sample may be too small . Do not add control solution to both sides

of the strip . The test strips are for single use only . Do not reuse test

strips . Discard used test strips .

English