Performing a Control Solution Test

1.

Check control solution expiration date .

2.

Remove test strip from vial .

3.

Insert test strip to turn meter on .

Note: The meter turns off after 2 minutes of inactivity .

Remove and reinsert the unused test strip to restart

the meter .

If the meter does not turn on go to Troubleshooting section .

IMPORTANT: Do not use control solution past expiration date . Discard

control solution 3 months after opening or on the expiration date

printed on the bottle whichever comes first . (Example: open April 15,

discard July 15; write the discard date on the side of the bottle .)