3.1.2

Humidity Sensor Specification

Sensor type:

Capacitative

Measuring range:

0 - 100% RH

Calibrated accuracy:

±3% RH

Hysteresis:

±1%

Linearity:

±1.5% RH between 5% - 95% RH

Time constant:

<10 s between 10% and 90% RH

Operating temperature: -25° to +80°C

3.1.3

Air Temperature & Humidity - Sensus Channel
Configurations

Description

Air temperature

Humidity

Units

o

C

%RH

Type

Analog

Analog

Information

Single ended
Gain x 1
No current source

Single ended
Gain x 1
No current source

Display Format

+##.#

###.#

Polynomial

-40, 100

0, 100

Enabled

Yes

Yes

Logged

Yes

Yes

Minimum Limit

-40.0

0.0

Maximum Limit

+60

100.0

3.2

Wind Speed Sensor

The sensitive three cup anemometer is
mounted at one end of the cross-arm. The
rotational speed is monitored using an
infra-red light source, interrupter and optical
sensor that provides 20 pulses from 0 to 5 V
per revolution. This non-contacting design
can measure wind speeds up to 75 m/sec.

The instrument has an anodised

aluminium alloy outer case that offers a
durable, all-weather construction suitable for
installation in harsh environments.

An output from this sensor is led to

a terminal box on the sensor cross arm, from
where a single sensor cable is taken to the
7-pin socket on the bottom panel of the environmental case.

3.2.1

Wind Speed Sensor Specification

Part number:

178031C

Sensor type:

Three cup anemometer

Transducer:

Optical interrupter

Maximum wind speed:

75 m/s

Starting velocity:

Typically 0.3 m/s

Distance constant:

Typically 3.5 m/s

Time constant:

Typically <0.5 s

Resolution:

7.84 cm

Non-linearity:

<±0.6 %

Accuracy:

±0.3 m/s below 3 m/s

Standard Sensors

01057

Wind speed sensor