CIRCUTOR CVM96 Series User Manual

Page 26

Advertising
background image

















----- Supply network analyzer CVM 96 ---------- ------- Page No. 25

a.- If an ENERGY parameter is chosen: kW.h (31), kvarh.L (32) or kvarh.C(33)

OUT 1

RELAY 1

PULS

xxxx

kW / pulse (1)

(1) Value of energy in kW : four digits with floating decimal point


- SETUP procedure:

- "max" key: to modify the value of the blinking value.
Every time it is pressed the value is increased.

- "min" key: to validate the blinking value and go to the next digit.

NOTE : When the last digit is reached, the position of the decimal can be move
point with the "max" key.

Example for setting a 500 W / 1 pulse:

Firstly we enter the value, 0500, and following we place the decimal point at

the right position with the "max" key 0.500 kW.