27.0 Troubleshooting guide

Trouble

Reason

Correction

Device doesn’t add
to connect software

Device authentication code incorrect.

Enter correct 4 charactor key code on rear of
product into software.

Weak signal and/or device is outside
connect stick range.

Move device closer to stick. Also refer to positioning
your devices document on www.bluelab.com

Device doesn’t
reconnect when I
move it to its location

Device outside connect stick range.

Move device closer to stick. Also refer to positioning
your devices document on www.bluelab.com

Device loses
connection when
it’s signal strength
is good

Device may be connected through
another device, if the closest device
drops out all connected devices also
drop out.

Increase the signal strength of the device
connected to the stick.

pH reading
inaccurate

pH probe not plugged in.

Connect pH probe. Check pH probe connection.

Using factory default calibration.

Calibrate pH probe.

Contaminated pH probe / glassware
not clean.

Clean pH probe, then calibrate.

Calibration old.

Calibrate pH probe.

Broken glass bulb, tube or connector.

Replace pH probe.

pH probe damaged or old.

Replace pH probe.

Bad grounding (noisy/jumpy pH
readings).

Earthlink required. Attach temperature probe. Go
to Menu, turn Earthlink on.

Display shows
‘Failed <e>’
during calibration

Old or contaminated solutions used for
calibration.

Use fresh calibration solutions.

Dirty or contaminated pH probe.

Clean pH probe.

pH probe tip been allowed to dry.

Hydrate pH probe.

pH probe damaged or old.

Replace pH probe.

No display

Mains not switched on.

Switch mains power on.

Power adaptor not plugged in.

Plug power adaptor into the socket marked ‘PWR’.

Mains on, power adaptor plugged in.

Replace power adaptor.

No display after
initial LCD test, when
plugged in

Backlight set to minimum.

Increase Backlight setting in MENU, or hold <cal>
button down while applying power, to restore
Factory Defaults.

pH displays
‘or’, ‘ur’, ‘-.-‘

‘or’ Over range pH.
‘ur’ Under range pH.

Check pH probe connection. pH probe could be
faulty. Clean pH probe, then calibrate.

‘-.-’ Temperature over/under range.

Solution <0°C / 32°F or >51°C / 122°F. Check
solution temperature. Ensure temperature probe
plug is fully inserted.

Temperature
displays
‘or’, ‘ur’, ‘- -‘

‘or’ Over range temperature.
‘ur’ Under range temperature.
‘- -’ Temperature probe not connected.

Solution >51 °C / 122 °F.
Solution <0 °C / 32 °F.
Ensure temperature probe plug is fully inserted.
Temperature probe is faulty, replace.

pH symbols fl ashing

It has been more than a month since
last pH calibration.

Clean pH probe, then calibrate.

‘HELP’ fl ashing
[Ineffective control
detected]

Dosing solution container empty.

Refi ll dosing solution container.

Dose direction setting incorrect for
your system.

Ensure setting refl ects dosing solution in use. i.e.
For Acid select Down.

Dose on/off times incorrect.

See section 18.0 for setting Dose On/Off times
correctly.

Output from tube not dripping into
solution.

Ensure tube output drips into tank.

Solution is not mixing.

Ensure pH correction solution is being mixed in
tank.

pH probe not seeing pH changes, after
dosing

Ensure pH probe is in tank/reservoir solution
with adequate mixing.