28.0 Frequently asked questions

Question

Answer

Why is the ‘’ or ‘’

fl ashing but the pump

is not turning?

The ‘

’ or ‘

’ will fl ash when ever the pH controller connect is in a dose

cycle. This includes the ‘OnTime’ and ‘OffTime’. The pump does not turn

during the ‘OffTime’.

What are dosing

lockouts?

A feature that stops the pH controller connect from dosing if:
• The temperature probe measures a solution temperature of less

than 0 °C / 32 °F or more than 50 °C / 122 °F.

• The pH reading does not change after 15 dose cycles.

Do I need to use the

temperature probe

with ATC for pH?

You can omit use of the ATC probe if:-
1) You do not use EarthLink On setting to stabilise the pH value
2) Your solution temperature is stable and you calibrate the pH probe

in calibration solutions at the same temperature as the reservoir /

tank solution.
3) Your pH is close to 7.0 pH

Why should I use

earthlink?

To remove/reduce any “mains earthloop” issues affecting the pH value’s

stability.

How do I use Earthlink?

Ensure the ATC probe is installed and in the same solution as the pH

probe.
Press ‘

’ to view status of pH mV value behaviour. Change the Earthlink

setting in MENU and watch the pH mV value again.
Select the Earthlink option that gives the least pH mV noise/variation.
If neither option improves pH stability:
1) Calibrate pH probe, see section 14.0.
2) Suspect Mains electrical problems. Obtain Electrician assistance.

How do I determine

the best Dose OnTime/

OffTime values?

Adjust OnTime such that 3 to 5 doses shifts the pH by only 0.1pH.
If <= 2 doses moves pH more than 0.1pH, you risk overdosing. (pH

changes past Required Value).
If it takes >=5 doses you will have a slow response to any changes. You

may also get “HELP” displayed if pH hasn’t moved enough if 15 dose

cycles.
If < 3 doses at 1 sec OnTime shifts the pH more than 0.1pH, you will need

to dilute the dosing solution to a lower concentration.

Adjust OffTime such that the last dose is fully mixed before the next

dose cycle starts.
If OffTime is too short, you risk overdosing (pH changes past Required

Value).
If OffTime is too long, correction to pH changes will take longer than

necessary.

How do I reset the pH

controller connect to

“as Shipped Defaults”?

Hold <cal> button down and apply power. Release button when “Restored

Factory Defaults” appears on screen.

How do I reset

pH calibration to

“Defaults”?

You shouldn’t need to. The pH probe can be calibrated anytime to the pH

controller connect. Follow the calibration steps in section 14.0.