2.4 Mounting

Install the CORI-FLOW in accordance to the direction of the FLOW arrow. The arrow for flow direction is

indicated on the instrument, between process fittings.

Mount the CORI-FLOW on a firm and steady base.

Avoid in any case heavy accelerations or mechanical shocks on the instrument.

For liquids, mount the CORI-FLOW at a level in the pipe line system where gas enclosure

is not possible and

be certain that the CORI-FLOW is filled all the time while operating.

For gases mount the CORI-FLOW at a level in the pipe line system where condensate can not collect in the

CORI-FLOW.

Verification of leaks is required prior starting up of the process.