For liquids, Calibration procedure, 2 for liquids – Bronkhorst CORI-FLOW User Manual

Page 24: 1000 p k, H ( kg, 6 calibration procedure

Advertising
background image

BRONKHORST CORI-TECH B.V.

4.5.2 For Liquids

Determine desired ∆p across valve.

p must be at least 50% of supply pressure

Kv-value calculation

1000

p

K

m

v

Φ

=

ρ

ρ


Units:

flow

-

)

/h

(kg

m

Φ

density

-

ρ

(kg / m ) at 20 C

3

and 1 atm.

delta p

-

p(bard)

The orifice bore diameter can be determined by:

d 7.6 K

v

=

[mm]

For C2 type of valves the diameter of the orifice can be calculated as shown above or looked up in the graph

below.

Φ

Φ

v 2

v

2

H O =

customer

customer

H O

ρ

ρ

in which:

Φ

v

= volume flow

ρ

= density

If liquids have viscosity’s >15 cs (water = 1 cs), the flat orifice and plunger type control mechanism cannot be

used. For measuring systems only check maximum possible viscosity with factory.

4.6 Calibration procedure

All instruments are factory calibrated. For re-calibration or re-ranging contact supplier or factory.