5 readout in mass flow or volume flow, 1 adjust instrument unit type, Readout in mass flow or volume flow – Bronkhorst CORI-FILL User Manual

Page 10: Adjust instrument unit type, 5readout in mass flow or volume flow

Advertising
background image

9.17.075

Bronkhorst Cori-Tech BV

10

5

Readout in mass flow or volume flow

5.1

Adjust instrument unit type

At the Tab “Basic” it is possible to prepare the instrument for mass flow readout, like e.g. g/h or kg/h or volumetric readout values

like e.g. l/min or ml/h. The counter will automatically switch to similar units like g or ml etc.

As the sensor can measure actual density, batch size can be given in mass and volume. To measure in volume, select “Volume

Flow”, the density should be specified manually.

When “Actual Volume Flow” is selected, the actual density will be used to convert mass flow to volume flow.

Normal Volume Flow and Standard Volume Flow are typical gas (mass)flow units. It will be necessary to enter the density at normal

(0 °C) and 1 atm. or at standard (20 °C) and 1 atm. conditions to be able to get the correct volumetric readout.

In order to change filter settings please see section 5.3.