On-off dosing using a shut-off Valve

Dosing example:

Counter limit = 10.00g.

Measured counter value = 9.98g

Counter value line in FlowPlot = 99.8%

Give a setpoint of any value, a shut-off valve will either be open or closed, so

setpoint value is irrelevant.

Please use a correct and stable inlet pressure

Because of the self learning principle (overrun correction) the requested

counter value will always be very accurate.

The Counter control gain option has no influence on dosing with a shut-off

valve. This formula is explained in section 8.

Overrun correction (on-off control):

A correction value between 0 and 1 can be entered. Therefore, when a value

is entered, the system will close the valve a fraction earlier or later,

depending on the overrun (or underrun). An overrun correction of 0.8 has

proven to be a proper functioning value. The closing value for the valve is

calculated by the following formula:

Closing value = Limit value – (Overrun correction *

(Last actual dosing value – Limit value)).

For example: if the last limit value for the dosing was 10g and the actual measured value was 10.50g, considering a overrun

correction factor of 0.8, for the next dosing the valve will close at: 10 – ( 0.8 * (10.5 – 10) ) = 10 – 0.40 = 9.60g

This correction is a continuous process, therefore it ensures that the requested dosing value will come to 10g.

Lower correction values can be used to filter instabilities in supply pressures, this will result in more cycles before the requested

batch is reached. Higher values result in fast correction and less cycles before final batch is reached.

PID settings can't be used in combination with shut-off valves!