8.2

Controller options

Controller options should be set for fast response. Below typical (fast response) settings for a pump controlled by a mini CORI-

FLOW instrument.

Valve powers, with the learn open function the instrument will

determine what the opening voltage of the used actuator is

(e.g. a valve or a pump).

After this value is known the opening response will be adapted to

decrease the dead time during setpoint steps from zero.

Setpoint slope. Normally = 0, except when PID-Td > 0.

PID-Kp: Typical = 10. Increase in range 10…100 to make controller

faster. Decrease to slow down.

PID-Ti: Typical = 0.1. Decrease in range 0.02..0.1 to make controller

faster. Increase from 0.1…2 to slow down.

PID-Td: Typical = 0. Increase a little (0…2 sec.) to differentiate dead

volume if necessary.

Special adjustment for PID-Kp at certain situations for extra speeding

up or slowing down.

i

www

For more detailed information about FlowPlot see:

FlowPlot manual

http://www.bronkhorst-cori-tech.com/files/downloads/manuals_english/917030manual_flowplot.pdf

Extra information

The PID scale can be adjusted by double clicking the highest value on the scale as shown below.