Fluke Biomedical QED 6 User Manual

Page 23

Advertising
background image

Introduction and Specifications

Specifications

1

1-11

Synchronization Measurements

Range ..................................................................... 0-199.9 ms
Measurement.......................................................... From peak of R-wave

From base of R-wave

Accuracy ................................................................. 1 % of fullscale or ±2 ms

ECG Waveforms

QRS complex

Rates .................................................................. 30, 60, 120, 180, 240 BPM
Rate Accuracy .................................................... ±1 % of setting
Amplitude............................................................ Fixed at 1 mV Lead II

(RA-LL)
Fixed at 1.1 mV
(Apex- Sternum)

Amplitude Accuracy ............................................ ±2 % (RA-LL)

±10 % (Apex-Sternum)

Performance Waveforms

Pulse....................................................................... 30, 60 BPM, pulse width 60 ms
Triangle Wave......................................................... 2 Hz
Square Wave ......................................................... 0.125 Hz, 2 Hz, 50 % duty cycle
Sine Waves............................................................. 10, 40, 50, 60, 100 Hz
Time Base Accuracy............................................... ±1 % of setting

Amplitude............................................................ Fixed at 1 mV Lead II

(RA-LL)
(Triangle wave 2 mV Lead II)
(RA-LL)
Fixed at 1.1 mV
(Apex-Sternum)

Amplitude Accuracy ............................................ ±2 % (RA-LL)

±10 % (Apex-Sternum)