Resetting the analyzer to factory defaults – Fluke Biomedical QED 6 User Manual
Page 61
![background image](/files/785496/content/doc061.png)
Operation
Resetting the Analyzer to Factory Defaults
2
2-35
PACER OUTPUT MEASUREMENTS
Pulse
Rate:
40
PPM
Pulse
Width:
0.0
ms
Peak
Current:
0.0
ma
Peak Volta
g
e:
0.0
V
Ave
Current
0.0
ma
Ave Volta
g
e:
0.0
V
PACER REFRACTORY MEASUREMENTS
Selected Pacin
g
Rate:
40 PPM
Pulsed Refractory Period:
329.2 ms
Sensed Refractory Period: 195.4 ms
fcf054.eps
Figure 2-11. Automatic Sequence Output with Header (cont.)
Resetting the Analyzer to Factory Defaults
To reset the Analyzer defaults:
1. From
Main Menu 2
, press the
AUTO
softkey to display the
Autosequences
menu:
fcf048.eps
2. Press the
DEFS
(defaults) softkey to access the
restore
menu.
fcf049.eps
3. Select one of the following options:
•
Press
(ALL)
to restore all 28 factory default programs.
•
Press
+
to scroll through the programs; stop when the desired
program name appears.