Resetting the analyzer to factory defaults – Fluke Biomedical QED 6 User Manual

Page 61

Advertising
background image

Operation

Resetting the Analyzer to Factory Defaults

2

2-35

PACER OUTPUT MEASUREMENTS
Pulse

Rate:

40

PPM

Pulse

Width:

0.0

ms

Peak

Current:

0.0

ma

Peak Volta

g

e:

0.0

V

Ave

Current

0.0

ma

Ave Volta

g

e:

0.0

V

PACER REFRACTORY MEASUREMENTS
Selected Pacin

g

Rate:

40 PPM

Pulsed Refractory Period:

329.2 ms

Sensed Refractory Period: 195.4 ms

fcf054.eps

Figure 2-11. Automatic Sequence Output with Header (cont.)

Resetting the Analyzer to Factory Defaults

To reset the Analyzer defaults:

1. From

Main Menu 2

, press the

AUTO

softkey to display the

Autosequences

menu:

fcf048.eps

2. Press the

DEFS

(defaults) softkey to access the

restore

menu.

fcf049.eps

3. Select one of the following options:

Press

(ALL)

to restore all 28 factory default programs.

Press

+

to scroll through the programs; stop when the desired

program name appears.