Operation

Evaluating Ability to Fire

2

2-13

OFF

DEFIB

Defibrillator/Pacer

PACER

QED-6

DEFIBRILLA

TOR A

NA

LYZER

STERNUM (-)

BAT

TE

RY / 9VD

C

RS-232 / 10101

ECG

V

C

LL

F

LA

L

RL

N

RA

R

DEFIB

APEX (+)

PA

CE

R

1000 j

100 j

SOFTKEYS

fcf010.eps

Figure 2-7. ECG Lead Configuration

3. From the

Energy

menu, press the

WAV>

softkey until the

VFIB

option

appears; then press the

VFIB

softkey.

A ventricular fibrillation waveform is simulated by the Analyzer through
the ECG jacks and paddle plates. When a discharge is complete, the
Analyzer outputs a 90 BPM Normal Sinus Rhythm.

4. Press the

WAV>

softkey until the

VTACH

option appears; then press the

VTACH

softkey.

A ventricular tachycardia is simulated by the Analyzer through the ECG
jacks and paddle plates. When the discharge is complete, the Analyzer
outputs a 90 BPM Normal Sinus Rhythm.