Operation pacemaker (non-invasive) testing – Fluke Biomedical QED 6 User Manual

Page 49

Advertising
background image

Operation

Pacemaker

(Non-Invasive)

Testing

2

2-23

QED-6

DEFIBRILLATOR ANALYZER

STERNUM (-)

BATTERY / 9VDC

RS-232 / 10101

ECG

V

C

LL

F

LA

L

RL

N

RA

R

DEFIB

APEX (+)

PACER

1000 j

100 j

SOFTKEYS

O

u

tp

u

t

Pacer

Adapters

Pacer

ECG Leads

RA

LA

LL

fcf015.eps

Figure 2-9. Connecting Pacemaker Output to Analyzer

Note

The Pacer can be in either demand or non-demand mode.

5. Press the

MEAS

softkey to display the following:

fcf031.eps

Three hyphens (---) indicate that no pacer pulses were received.

6. Set the pacer at various current and heart rate settings. The results are

displayed. Press the

PRINT

softkey for a hard copy.