SSi Manual 4549

Page 10 of 16

3-Gas Analyzer PGA3000

Revision Level E

About/Sign-On – Menu Page 11

This page is the sign on
screen that shows the

SSi logo, address, and
phone number. Also

shown are the firmware

version number, the
unit serial number, and

the date of the last
factory calibration.

Some units will also

show the internal
temperature of the unit.

Start logger – Menu Page 12

This page is used to

name and store a data
log file. The PGA3000 is

capable of storing 10

tests (files) of 500
minutes each. To start

storing data, first select
a test number, 0 – 9.

This specifies the

storage location (file) of
the data. Then input a

name and then an ID number. The name selected from the list at the bottom of
the page by number (0 – 9). The ID number is any value from 0 to 999. Then

cursor down to the Stop/Run then select and press the A key to start or stop

storing data. STORING DATA will appear near the bottom of the screen
whenever the data logging is active. Only one test number can be active at a

time. If the test number is changed and started, then the previous one is
stopped. When a test is started, all previous data in that test is cleared. See

Basic Operating Theory (page 4 of 15) for keypad key assignments.