Super Systems PGA3000 User Manual

Page 7

Advertising
background image

SSi Manual 4549

Page 7 of 16

3-Gas Analyzer PGA3000

Revision Level E

IR Status Display – Menu Page 1

The IR Status Display

shows the current

readings of CO, CO2
and CH4. Also shown

is the relative flow rate
of the sample by a fuel

gauge on the left-hand

side of the screen.
The sample pump may

be turned on or off by
pressing the A key.

IR Overall Status Display – Menu Page 2

The IR Overall Status

Display shows measured

values and calculated
values, and allows entry

of data for the
calculations. The

operator can enter the

furnace temperature,
probe millivolts, probe

temperature, and either
the CO Factor or the

Process Factor. The probe calculations will be based on which value was entered

last (i.e. COF or PF). An * after the data values indicates which value is used in
the calculations. The equivalent value of the other value is computed and

displayed. For example if the probe millivolts is 1100, the probe temperature
1650 and the COF is 200, then the probe %C is calculated as 0.51% and the

equivalent PF is calculated as 143. From the three gases and the furnace

temperature the IR %C and equivalent millivolts is computed. The IR %C and
the probe %C are used to compute suggested COF and PF values. Data values

are changed by using the cursor keys (B and C) to highlight the value and then
entering the new value followed by the A key. Suggested COF or PF can be

entered in your control instrumentation so that the %C calculated from the

carbon probe readings will be the same as the %C calculated by the IR 3-Gas
readings.

WARNING – Large changes should be verified by shim stock analysis.

Advertising