Hydrogen (H

2

) Sensor Operations Manual

Super Systems Inc.

Page 10 of 17

Figure 9 - Overview screen

Sensor Output Configuration

The Configuration screen (Figure 10) allows basic configuration of the H

2

cell.

Figure 10 - Configuration screen

The Instrument Mode is the primary PV (Process Variable) setup. This is mainly used by the

touch screen and normally does not need to be changed as the % H

2

is always available. The

output sources are the PVs that will be retransmitted via the selected output. The Output Zero

is the PV value that will result in a 4 mA output and the output span value is the PV value that

will result in a 20 mA output.

Sensor Output Calibration

Output calibration calibrates the outputs. Each output can be zeroed and spanned. To perform

this calibration a reliable measurement device capable of accurately measuring a 4-20mA

signal will be required. An example screen is shown in Figure 11; a screen showing the success

of an output calibration can be seen in Figure 12.